Professional 98mm X 98mm X 111mm Online PH Meter For Mining And Smelting Specifications :
Price | Contact Us |
Payment Terms | T/T, L/C, Western Union, etc. |
Delivery Detail | 10-25 working days. |
Professional 98mm X 98mm X 111mm Online PH Meter For Mining And Smelting Detail Introduction :
Description:
A new generation of high-end microcomputer-based instruments with LCD display, Chinese and English menu operation, full intelligence, multi-function, high measurement performance, strong environmental adaptability, etc., can accurately measure the pH value and temperature of the solution.
Main features:
LCD display, Chinese and English menu, fast switching, easy to operate
High-precision triple composite electrode, accurate and convenient measurement
With manual and automatic temperature compensation function
Advanced microcontroller technology, high performance, low power consumption
The motherboard adopts SMD technology and highly integrated isolation circuit design, with good anti-interference performance, stability and durability
Standard output signal type is optional, two sets of relay control switch alarm can be set arbitrarily With storage function, users can set and store measurement data according to their needs. Convenient for query
The historical curve can be displayed, and the PH change of water quality is convenient for query.
Specification:
Display: 128X64 dot matrix LCD
Range: (0.00-14.00) pH Indication error: ± 0.02pH
Resolution: 0.01 pH
Input impedance: >1X1012Ω
Repeatability: not more than 1%
Temperature sensor: Pt1000 Temperature measurement range: (0.0-99.9)℃
Temperature compensation range: (0.1-60.0)℃
Water sample temperature: (5-60) ℃ Ambient temperature: (5-45) ℃
Environmental humidity: not more than 90%RH (no condensation)
Storage and transportation temperature: (-25-55)℃ Power supply: AC (220 ± 22) V, frequency (50 ± 1) HZ, DC 24 VDC,
Dimensions: 98mm X 98mm X 111mm (length X width X height)
Hole size: 96 mm X 96 mm Weight: 1.0kg Galvanic isolation output: 0-10mA, 0-20mA, 4-20mA optional
Alarm relay: 2 normally open points can be set arbitrarily, AC220V 3A /DC30V 3A
Protection level: IP65, power-off storage: >10 years
Secondary instrument installation method: open-hole/wall-mounted/rack-mounted
Electrode installation method: flow-through type / sinking type / flange type / pipeline type